Skip to main navigation Skip to search Skip to main content

Axial and radial diffusion coefficients in a liquid chromatography column and bed heterogeneity

Research output: Contribution to journalArticle

36 Citations (Scopus)

Abstract

The axial and transverse diffusion coefficients of a band of iodine in a chromatographic column were measured optically as a function of time. It was found that the axial diffusion coefficient remains constant even when the edges of the sample band get close to the wall. By contrast, the radial diffusion coefficient decreases progressively with increasing time when the edges of the sample band leave the core region and begin to diffuse inside the wall region. The local axial and transverse diffusion coefficients of the band decrease from the column center toward the wall. Hence, the increase in local height equivalent to a theoretical plate observed in the region close to the wall must be explained by increasing mass transfer resistances and degree of heterogeneity of the bed.
Original languageEnglish
Pages (from-to)1-12
Number of pages12
JournalJournal of Chromatography A
Volume994
Issue number1-2
Publication statusPublished - 2003

Keywords

  • Chromatographic analysis
  • Diffusion
  • Iodine
  • Packings (Chromatography)
  • Stationary phase (Chromatography)
  • Packed columns
  • Stationary phases, LC
  • Packing density
  • Diffusion coefficients

Fingerprint

Dive into the research topics of 'Axial and radial diffusion coefficients in a liquid chromatography column and bed heterogeneity'. Together they form a unique fingerprint.