Characterisation of electroosmotic flow in capillary electrochromatography columns

H. S. Dearie, V. Spikmans, N. W. Smith, F. Moffatt, S. A.C. Wren, K. P. Evans

Research output: Contribution to journalArticlepeer-review

24 Citations (Scopus)

Abstract

Currently available capillary electrochromatography (CEC) instrumentation using UV-Vis detection dictates the use of duplex columns. Due to discontinuities (electric field strength and conductivity) that arise at the boundary between the packed and open sections in these columns, the determination of the electroosmotic flow (EOF) is complicated. Thiourea has been found to be an accurate EOF marker under the conditions employed in this study. By injecting this compound onto a fully packed column and comparing the obtained mobilities with those calculated from measured zeta potential values a value for tortuosity has been obtained. The use of laser Doppler velocimetry (LDV) for the measurement of zeta potential has been found to be the most direct and rapid method of characterising silica support materials in terms of electroosmotic mobility. The open section in duplex CEC columns has been shown to influence the actual column flow-rate. The EOF measured using duplex columns of varying packed and open section lengths have been compared with those obtained for a fully packed column.

Original languageEnglish
Pages (from-to)123-131
Number of pages9
JournalJournal of Chromatography. A
Volume929
Issue number1-2
DOIs
Publication statusPublished - 21 Sept 2001
Externally publishedYes

Keywords

  • Electrochromatography
  • Electroosmotic flow
  • Tortuosity
  • Zeta potential

Fingerprint

Dive into the research topics of 'Characterisation of electroosmotic flow in capillary electrochromatography columns'. Together they form a unique fingerprint.

Cite this