Code-division-multiplexed electrical impedance tomography spectroscopy

Research output: Contribution to journalArticle

20 Citations (Scopus)

Abstract

Electrical impedance tomography uses multiple impedance measurements to image the internal conductivity of an object, such as the human body. Code-division multiplexing is proposed as a new method that can provide simultaneous impedance measurements of the multiple channels. Code division provides clear advantages of a wide frequency range at reduced cost and reduced complexity of sources. A potential drawback is the lack of perfectly orthogonal code sets. This caused an increase of 0.62% in root-mean-square spectral error when two codes were used to record two impedance channels simultaneously on a low-pass filter network. The method described provides images and spectra which are equivalent to the conventional time-multiplexed method, with increases in frequency resolution and measurement speed which may be of benefit in some applications of electrical impedance tomography spectroscopy.
Original languageEnglish
Pages (from-to)332-338
Number of pages7
JournalIEEE Transactions on Biomedical Circuits and Systems
Volume3
DOIs
Publication statusPublished - 2009

Keywords

  • electrical impedance tomography
  • multiplexing

Fingerprint

Dive into the research topics of 'Code-division-multiplexed electrical impedance tomography spectroscopy'. Together they form a unique fingerprint.

Cite this