TY - JOUR
T1 - Comparison of electromyography and kinemyography during recovery from non-depolarising neuromuscular blockade
AU - Stewart, P. A.
AU - Freelander, N.
AU - Liang, S.
AU - Heller, G.
AU - Phillips, S.
PY - 2014
Y1 - 2014
N2 - In this study, two commercially available quantitative neuromuscular function monitoring techniques, electromyography (EMG) and kinemyography (KMG), were compared with respect to repeatability and accuracy during late recovery from neuromuscular blockade. Train-of-four (TOF) ratios were recorded in 30 patients using KMG and EMG at the adductor pollicis muscle. Measurements were taken on the same hand using the Datex-Ohmeda NeuroMuscular Transmission monitor (GE Healthcare, Helsinki, Finland). Instrumental precision was evaluated using the coefficient of repeatability, while accuracy was assessed using the bias and limits of agreement. The coefficients of repeatability were similar for both techniques (0.035 for KMG and 0.043 for EMG), indicating a similar level of precision. KMG overestimated the TOF ratios measured with EMG with a bias of 0.11 (95% limits of agreement: -0.13 to 0.35). At a TOF ratio of 0.90 the bias was 0.08 (95% limits of agreement: -0.08 to 0.25). This means that at a TOF ratio of 0.90 measured with KMG will be approximately equivalent to a TOF ratio of 0.80 measured with EMG at the adductor pollicis muscle, but it may indeed be as low as 0.65 or as high as 1.00. Therefore, TOF ratios measured by KMG and EMG cannot be used interchangeably.
AB - In this study, two commercially available quantitative neuromuscular function monitoring techniques, electromyography (EMG) and kinemyography (KMG), were compared with respect to repeatability and accuracy during late recovery from neuromuscular blockade. Train-of-four (TOF) ratios were recorded in 30 patients using KMG and EMG at the adductor pollicis muscle. Measurements were taken on the same hand using the Datex-Ohmeda NeuroMuscular Transmission monitor (GE Healthcare, Helsinki, Finland). Instrumental precision was evaluated using the coefficient of repeatability, while accuracy was assessed using the bias and limits of agreement. The coefficients of repeatability were similar for both techniques (0.035 for KMG and 0.043 for EMG), indicating a similar level of precision. KMG overestimated the TOF ratios measured with EMG with a bias of 0.11 (95% limits of agreement: -0.13 to 0.35). At a TOF ratio of 0.90 the bias was 0.08 (95% limits of agreement: -0.08 to 0.25). This means that at a TOF ratio of 0.90 measured with KMG will be approximately equivalent to a TOF ratio of 0.80 measured with EMG at the adductor pollicis muscle, but it may indeed be as low as 0.65 or as high as 1.00. Therefore, TOF ratios measured by KMG and EMG cannot be used interchangeably.
KW - anesthesia
KW - electromyography
UR - http://handle.uws.edu.au:8081/1959.7/uws:37391
UR - http://search.proquest.com/docview/1526026194?accountid=36155
M3 - Article
SN - 0310-057X
VL - 42
SP - 378
EP - 384
JO - Anaesthesia and Intensive Care
JF - Anaesthesia and Intensive Care
IS - 3
ER -