TY - JOUR
T1 - Comparison of kinemyography and electromyography during spontaneous recovery from non-depolarising neuromuscular blockade
AU - Khandkar, C.
AU - Liang, S.
AU - Phillips, S.
AU - Lee, C. Y.
AU - Stewart, P. A.
PY - 2016
Y1 - 2016
N2 - This study compared two commercially available quantitative neuromuscular function monitoring techniques, kinemyography (KMG) and electromyography (EMG), to assess whether KMG could be used interchangeably with EMG to exclude residual neuromuscular blockade (RNMB). Train-of-four (TOF) ratios were recorded every 20 seconds using KMG at the adductor pollicis and EMG at the first dorsal interosseous of the same hand during spontaneous recovery from shallow neuromuscular blockade. TOF ratios were compared using Bland-Altman analysis for repeated measurements. The precision of each device was assessed by the repeatability coefficient. Agreement between devices was assessed by the bias and limits of agreement. Clinically acceptable agreement was defined as a bias <0.025 within limits of agreement ±0.05. We recorded 629 sets of TOF ratios from 23 patients. The repeatability coefficient for KMG was 0.05 (95% confidence interval [CI] 0.05 to 0.06) and for EMG 0.10 (95% CI 0.10 to 0.11). Overall, the bias of KMG TOF ratios against EMG TOF ratios was 0.11 (95% CI 0.10 to 0.12), with limits of agreement -0.11 to 0.32. In the 0.80 to 0.99 TOF range, the bias was 0.08 (95% CI 0.06 to 0.09) and the limits of agreement were -0.12 to 0.27. Overall, TOF ratios measured by KMG were on average 0.11 higher than EMG. In the 0.80 to 0.99 TOF range, KMG TOF ratios were 0.08 higher. EMG and KMG are not interchangeable because the bias is large and the limits of agreement are wide. Thus a maximum TOF ratio of 1.0 on KMG may not exclude RNMB.
AB - This study compared two commercially available quantitative neuromuscular function monitoring techniques, kinemyography (KMG) and electromyography (EMG), to assess whether KMG could be used interchangeably with EMG to exclude residual neuromuscular blockade (RNMB). Train-of-four (TOF) ratios were recorded every 20 seconds using KMG at the adductor pollicis and EMG at the first dorsal interosseous of the same hand during spontaneous recovery from shallow neuromuscular blockade. TOF ratios were compared using Bland-Altman analysis for repeated measurements. The precision of each device was assessed by the repeatability coefficient. Agreement between devices was assessed by the bias and limits of agreement. Clinically acceptable agreement was defined as a bias <0.025 within limits of agreement ±0.05. We recorded 629 sets of TOF ratios from 23 patients. The repeatability coefficient for KMG was 0.05 (95% confidence interval [CI] 0.05 to 0.06) and for EMG 0.10 (95% CI 0.10 to 0.11). Overall, the bias of KMG TOF ratios against EMG TOF ratios was 0.11 (95% CI 0.10 to 0.12), with limits of agreement -0.11 to 0.32. In the 0.80 to 0.99 TOF range, the bias was 0.08 (95% CI 0.06 to 0.09) and the limits of agreement were -0.12 to 0.27. Overall, TOF ratios measured by KMG were on average 0.11 higher than EMG. In the 0.80 to 0.99 TOF range, KMG TOF ratios were 0.08 higher. EMG and KMG are not interchangeable because the bias is large and the limits of agreement are wide. Thus a maximum TOF ratio of 1.0 on KMG may not exclude RNMB.
KW - electromyography
KW - neuromuscular blocking agents
KW - neuromuscular transmission
UR - http://handle.uws.edu.au:8081/1959.7/uws:37918
UR - http://search.ebscohost.com/login.aspx?direct=true&db=a9h&AN=119373113&site=ehost-live&scope=site
M3 - Article
SN - 1448-0271
SN - 0310-057X
VL - 44
SP - 745
EP - 751
JO - Anaesthesia and Intensive Care
JF - Anaesthesia and Intensive Care
IS - 6
ER -