Comparison of setup accuracy of three different image assessment methods for tangential breast radiotherapy

Vikneswary Batumalai, Penny Phan, Callie Choong, Lois Holloway, Geoff P. Delaney

Research output: Contribution to journalArticlepeer-review

Abstract

Introduction: To compare the differences in setup errors measured with electronic portal image (EPI) and cone-beam computed tomography (CBCT) in patients undergoing tangential breast radiotherapy (RT). Relationship between setup errors, body mass index (BMI) and breast size was assessed. Methods: Twenty-five patients undergoing postoperative RT to the breast were consented for this study. Weekly CBCT scans were acquired and retrospectively registered to the planning CT in three dimensions, first using bony anatomy for bony registration (CBCT-B) and again using breast tissue outline for soft tissue registration (CBCT-S). Digitally reconstructed radiographs (DRR) generated from CBCT to simulate EPI were compared to the planning DRR using bony anatomy in the V (parallel to the cranio-caudal axis) and U (perpendicular to V) planes. The systematic (Σ) and random (σ) errors were calculated and correlated with BMI and breast size. Results: The systematic and random errors for EPI (ΣV = 3.7 mm, ΣU = 2.8 mm and σV = 2.9 mm, σU = 2.5) and CBCT-B (ΣV = 3.5 mm, ΣU = 3.4 mm and σV = 2.8 mm, σU = 2.8) were of similar magnitude in the V and U planes. Similarly, the differences in setup errors for CBCT-B and CBCT-S in three dimensions were less than 1 mm. Only CBCT-S setup error correlated with BMI and breast size. Conclusions: CBCT and EPI show insignificant variation in their ability to detect setup error. These findings suggest no significant differences that would make one modality considered superior over the other and EPI should remain the standard of care for most patients. However, there is a correlation with breast size, BMI and setup error as detected by CBCT-S, justifying the use of CBCT-S for larger patients.
Original languageEnglish
Pages (from-to)224-231
Number of pages8
JournalJournal of Medical Radiation Sciences
Volume63
Issue number4
DOIs
Publication statusPublished - 2016

Open Access - Access Right Statement

© 2016 The Authors. Journal of Medical Radiation Sciences published by John Wiley & Sons Australia, Ltd on behalf of Australian Society of Medical Imaging and Radiation Therapy and New Zealand Institute of Medical Radiation Technology. This is an open access article under the terms of the Creative Commons Attribution-NonCommercial-NoDerivs License, which permits use and distribution in any medium, provided the original work is properly cited, the use is non-commercial and no modifications or adaptations are made.

Keywords

  • breast
  • cancer
  • errors
  • radiotherapy
  • tomography

Fingerprint

Dive into the research topics of 'Comparison of setup accuracy of three different image assessment methods for tangential breast radiotherapy'. Together they form a unique fingerprint.

Cite this