TY - JOUR
T1 - Comparison of three reconstruction algorithms for low-dose phase-contrast computed tomography of the breast with synchrotron radiation
AU - Donato, Sandro
AU - Caputo, Simone
AU - Brombal, Luca
AU - Golosio, Bruno
AU - Longo, Renata
AU - Tromba, Giuliana
AU - Agostino, Raffaele G.
AU - Greco, Gianluigi
AU - Arhatari, Benedicta
AU - Hall, Chris
AU - Maksimenko, Anton
AU - Hausermann, Daniel
AU - Lockie, Darren
AU - Fox, Jane
AU - Kumar, Beena
AU - Lewis, Sarah
AU - Brennan, Patrick C.
AU - Quiney, Harry M.
AU - Taba, Seyedamir T.
AU - Gureyev, Timur E.
PY - 2025/7
Y1 - 2025/7
N2 - Background: Phase-contrast breast CT imaging holds promise for improved diagnostic accuracy, but an optimal reconstruction algorithm must balance objective image quality metrics with subjective radiologist preferences. Purpose: This study systematically compares three reconstruction algorithms—filtered back projection (FBP), unified tomographic reconstruction (UTR), and customized simultaneous algebraic reconstruction technique (cSART)—to identify the most suitable approach for phase-contrast breast CT imaging. Methods: Fresh mastectomy samples were scanned at the Australian synchrotron using monochromatic 32 keV X-rays, a mean glandular dose of 2 mGy, flat-panel detectors with 0.1 mm pixels, and 6-m distance between the rotation stage and the detector. Paganin's phase retrieval method was used in conjunction with all three CT reconstruction algorithms. Objective metrics, including spatial resolution, contrast, signal-to-noise, and contrast-to-noise, were evaluated alongside subjective assessments by seven experienced radiologists. Ratings included perceptible contrast, sharpness, noise, calcification visibility, and overall quality. Results: cSART excelled in objective metrics, outperforming UTR and FBP. However, subjective evaluations favored FBP due to its higher image contrast, revealing a discrepancy between objective and subjective assessments. Conclusions: The findings highlight the contrast-focused nature of radiologists’ subjective assessments and the potential of cSART for delivering superior objective image quality. These insights inform the development of hybrid evaluation tools and guide clinical translation for future live patient imaging studies.
AB - Background: Phase-contrast breast CT imaging holds promise for improved diagnostic accuracy, but an optimal reconstruction algorithm must balance objective image quality metrics with subjective radiologist preferences. Purpose: This study systematically compares three reconstruction algorithms—filtered back projection (FBP), unified tomographic reconstruction (UTR), and customized simultaneous algebraic reconstruction technique (cSART)—to identify the most suitable approach for phase-contrast breast CT imaging. Methods: Fresh mastectomy samples were scanned at the Australian synchrotron using monochromatic 32 keV X-rays, a mean glandular dose of 2 mGy, flat-panel detectors with 0.1 mm pixels, and 6-m distance between the rotation stage and the detector. Paganin's phase retrieval method was used in conjunction with all three CT reconstruction algorithms. Objective metrics, including spatial resolution, contrast, signal-to-noise, and contrast-to-noise, were evaluated alongside subjective assessments by seven experienced radiologists. Ratings included perceptible contrast, sharpness, noise, calcification visibility, and overall quality. Results: cSART excelled in objective metrics, outperforming UTR and FBP. However, subjective evaluations favored FBP due to its higher image contrast, revealing a discrepancy between objective and subjective assessments. Conclusions: The findings highlight the contrast-focused nature of radiologists’ subjective assessments and the potential of cSART for delivering superior objective image quality. These insights inform the development of hybrid evaluation tools and guide clinical translation for future live patient imaging studies.
KW - breast CT
KW - reconstruction algorithms
KW - synchrotron radiation
UR - http://www.scopus.com/inward/record.url?scp=105011169244&partnerID=8YFLogxK
U2 - 10.1002/mp.17950
DO - 10.1002/mp.17950
M3 - Article
AN - SCOPUS:105011169244
SN - 0094-2405
VL - 52
JO - Medical Physics
JF - Medical Physics
IS - 7
M1 - e17950
ER -