TY - JOUR
T1 - Exposure (mAs) optimisation of a multi-detector CT protocol for hepatic lesion detection : are thinner slices better?
AU - Dobeli, K. L.
AU - Lewis, Sarah Jane
AU - Meikle, S. R.
AU - Thiele, D. L.
AU - Brennan, P. C.
PY - 2014
Y1 - 2014
N2 - Introduction The purpose of this work was to determine the exposure-optimised slice thickness for hepatic lesion detection with CT. Methods A phantom containing spheres (diameter 9.5, 4.8 and 2.4 mm) with CT density 10 HU below the background (50 HU) was scanned at 125, 100, 75 and 50 mAs. Data were reconstructed at 5-, 3- and 1-mm slice thicknesses. Noise, contrast-to-noise ratio (CNR), area under the curve (AUC) as calculated using receiver operating characteristic analysis and sensitivity representing lesion detection were calculated and compared. Results Compared with the 125 mAs/5 mm slice thickness setting, significant reductions in AUC were found for 75 mAs (P < 0.01) and 50 mAs (P < 0.05) at 1- and 3-mm thicknesses, respectively; sensitivity for the 9.5-mm sphere was significantly reduced for 75 (P < 0.05) and 50 mAs (P < 0.01) at 1-mm thickness; sensitivity for the 4.8-mm sphere was significantly lower for 100, 75 and 50 mAs at all three slice thicknesses (P < 0.05). The 2.4-mm sphere was rarely detected. At each slice thickness, noise at 100, 75 and 50 mAs exposures was approximately 10, 30 and 50% higher, respectively, than that at 125 mAs exposure. CNRs decreased in an irregular manner with reductions in exposure and slice thickness. Conclusion This study demonstrated no advantage to using slices below 5 mm thickness, and consequently thinner slices are not necessarily better.
AB - Introduction The purpose of this work was to determine the exposure-optimised slice thickness for hepatic lesion detection with CT. Methods A phantom containing spheres (diameter 9.5, 4.8 and 2.4 mm) with CT density 10 HU below the background (50 HU) was scanned at 125, 100, 75 and 50 mAs. Data were reconstructed at 5-, 3- and 1-mm slice thicknesses. Noise, contrast-to-noise ratio (CNR), area under the curve (AUC) as calculated using receiver operating characteristic analysis and sensitivity representing lesion detection were calculated and compared. Results Compared with the 125 mAs/5 mm slice thickness setting, significant reductions in AUC were found for 75 mAs (P < 0.01) and 50 mAs (P < 0.05) at 1- and 3-mm thicknesses, respectively; sensitivity for the 9.5-mm sphere was significantly reduced for 75 (P < 0.05) and 50 mAs (P < 0.01) at 1-mm thickness; sensitivity for the 4.8-mm sphere was significantly lower for 100, 75 and 50 mAs at all three slice thicknesses (P < 0.05). The 2.4-mm sphere was rarely detected. At each slice thickness, noise at 100, 75 and 50 mAs exposures was approximately 10, 30 and 50% higher, respectively, than that at 125 mAs exposure. CNRs decreased in an irregular manner with reductions in exposure and slice thickness. Conclusion This study demonstrated no advantage to using slices below 5 mm thickness, and consequently thinner slices are not necessarily better.
UR - https://hdl.handle.net/1959.7/uws:74448
U2 - 10.1111/1754-9485.12104
DO - 10.1111/1754-9485.12104
M3 - Article
SN - 1754-9485
SN - 1754-9477
VL - 58
SP - 137
EP - 143
JO - Journal of Medical Imaging and Radiation Oncology
JF - Journal of Medical Imaging and Radiation Oncology
IS - 2
ER -