The use of parallel segmented outlet flow columns for enhanced mass spectral sensitivity at high chromatographic flow rates

  • Michelle Camenzuli
  • , Tiffany A. Goodie
  • , Danielle N. Bassanese
  • , Paul S. Francis
  • , Neil .W. Barnett
  • , Harald Ritchie
  • , James LaDine
  • , R. Andrew Shalliker
  • , Xavier A. Conlan

Research output: Contribution to journalArticlepeer-review

19 Citations (Scopus)

Abstract

RATIONALE Speed of analysis is a significant limitation to current high-performance liquid chromatography/mass spectrometry (HPLC/MS) and ultra-high-pressure liquid chromatography (UHPLC)/MS systems. The flow rate limitations of MS detection require a compromise in the chromatographic flow rate, which in turn reduces throughput, and when using modern columns, a reduction in separation efficiency. Commonly, this restriction is combated through the post-column splitting of flow prior to entry into the mass spectrometer. However, this results in a loss of sensitivity and a loss in efficiency due to the post-extra column dead volume. METHODS A new chromatographic column format known as 'parallel segmented flow' involves the splitting of eluent flow within the column outlet end fitting, and in this study we present its application on a HPLC electrospray ionization time-of-flight mass spectrometer. RESULTS Using parallel segmented flow, column flow rates as high as 2.5 mL/min were employed in the analysis of amino acids without post-column splitting to the mass spectrometer. Furthermore, when parallel segmented flow chromatography columns were employed, the sensitivity was more than twice that of conventional systems with post-column splitting when the same volume of mobile phase was passed through the detector. CONCLUSIONS These finding suggest that this type of column technology will particularly enhance the capabilities of modern LC/MS enabling both high-throughput and sensitive mass spectral detection.
Original languageEnglish
Pages (from-to)943-949
Number of pages7
JournalRapid Communications in Mass Spectrometry
Volume26
Issue number8
DOIs
Publication statusPublished - 2012

Keywords

  • amino acids
  • chromatography
  • electrospray ionization
  • high pressure liquid
  • instrumentation
  • mass spectrometry

Fingerprint

Dive into the research topics of 'The use of parallel segmented outlet flow columns for enhanced mass spectral sensitivity at high chromatographic flow rates'. Together they form a unique fingerprint.

Cite this